To ensure the integrity of your filters, it is crucial to conduct tests before and after operation. For hydrophilic filters, there are several methods available:
Both of these methods necessitate the use of costly equipment, such as a mass flow meter or an automated integrity tester, and a more intricate system installation if the test is to be conducted with filter in place. A budget-friendly approach is pressure decay testing.
Forward flow air diffusion measures the flow of a gas as it diffuses through a wetted membrane at a sustained pressure below the bubble point. It quantifies the gas flow required to maintain the test pressure on the upstream side as it traverses the membrane.
Conversely, the pressure decay test ascertains the pressure loss incurred due to this diffusion over a specified time duration. As air transitions from the upstream side to the downstream side, the upstream pressure will decrease in proportion to the upstream volume and the rate of air flow.
A typical setup for conducting a pressure decay test is depicted in Figure 1 (cartridge housed within a housing) and Figure 2 (Capsule). To ascertain the acceptable pressure loss, it is imperative to determine the upstream volume of the filter housing and lines after the air shutoff valve (V1).
Either of these configurations could be easily set up to operate on the filter without the need to remove it from the line. This is particularly crucial when conducting pre-use integrity checks after sanitization and/or post-use integrity checks to ensure that no alterations have occurred.
Figure 1: Cartridge Pressure Decay Test Attachment to Housing Vent
Figure 2: Capsule Pressure Decay Testing
Once the upstream volume is determined, the pressure decay limit (DP in the equations below) can be calculated from the following equation:.
(D) x (T) x (Pa) = (DP) x (Vh)
DP = (D) x (T) x (Pa) / (Vh)
Where:
D = Diffusion rate cc/min (max value from the product data sheet)
T = Time (minutes)
Pa = Atmospheric Pressure (14.7 psi)
DP = Pressure Drop (psi)
Vh = Upstream volume of apparatus (cc)
For help determining the test values or additional information, contact customerservice@criticalprocess.com or call us at (603) 880-4420.
For more information on Pressure Decay Testing, see:
Capsule Pressure Decay Testing
Cartridge Pressure Decay Testing